lyzer has been zeroed with a gas that contains a significant concen-
tration of O
2
(Figure 2-3).
•
0-1V Range Identifier
This 0 to 1 volt output represents each range with a particular
voltage as shown in Table 2-1.
•
Isolated 4-20 mA Signal (Oxygen Measurement)
This is a 4 to 20 mA output representing 0 to 100% of the scale,
with 4 mA equal to 0%, and 20 mA equal to 100% of that range.
This output may also range lower than 4 mA, especially if the
analyzer has been zeroed with a gas that contains a significant
concentration of O
2
(Figure 2-5).
•
Isolated 4-20 mA Range Identifier
This 4 to 20 mA output identifies individual ranges with discrete
current output as shown in Table 2-1.
Identifier
Identifier
Range
Voltage (V) Current (mA)
50 PPB
0.0
4.0
100 PPB
0.2
7.2
1 PPM
0.4
10.4
10 PPM
0.6
13.6
30 PPM
0.8
16.8
Table 2-1. Range Identifier
Figure 2-5. Analog Signal Output Offset
- 1/4
Full Scale
Value
- 1/4
Full
Scale
Value
-0.25
1 V
(0,0)
(0,0)
Full
Scale
Full
Scale
20 mA
Voltage (V)
Current (I)
O2
O2