Operations – Teledyne 3060e - Ultra Trace oxygen analyzer User Manual
Page 89

Teledyne Electronic Technologies
Analytical Instruments
4-41
Operations
Operations
Operations
Operations
Operations
Ultra
Ultra
Ultra
Ultra
Ultra T
T
T
T
Trace Oxyg
race Oxyg
race Oxyg
race Oxyg
race Oxygen Anal
en Anal
en Anal
en Anal
en Analyz
yz
yz
yz
yzer
er
er
er
er
6. When you make your selection, the
alarm number is displayed in the second
line. The alarm parameters are listed on
the screen. Use the
UP
and
DOWN
keys
to scroll through the values within each
alarm parameter. Press
SELEC
to select
the desired parameters. Press
ACCPT
to
enter all of the selected values into the
computer. These are now the set values
for this particular alarm.
•
Fail-safe option YES de-activates the alarm relay on the rear panel
(TB1, TB2, TB3, TB4 or TB5 for alarms 1 through 5) when an
alarm condition exists, including system power failure. If NO is
chosen, the relay will allow activation of the alarms during alarm
status.
Note that an external alarm in series with a power supply should be
connected across the C and NC (normally closed) relay contacts in
fail-safe mode, and across the C and NO (normally open) contacts in
non-fail-safe mode.
•
Level indicates the oxygen level that will activate the alarm.
•
High/Low indicates whether the level is a high- or low-threshold.
•
Latching option YES means that the alarm will continue to ring
after being triggered by an alarm status, even though the oxygen has
returned to acceptable levels. The ALRMR button on the keyboard
in the Analyze mode menu must be pressed to end the alarm.
•
Defeat option YES renders the alarm inoperative.
7. When you press
ACCPT
, the next screen
allows you to select another alarm to set
up. You can select another alarm (up to
5 alarms), or press
ESC
twice to return
to the Analyze mode.
8. Pressing
ESC
twice returns you to the
Analyze mode main menu. The appro-
priate alarm number and type flashes in
the upper right-hand corner of the
display if the alarm is tripped.
VALVES: ANALYZE C=9:59
T=25C
ALARM 1 SETUP
LEVEL = 1.00 PPM RELAY =TB1
HIG H/LO W HI FAILSAFE YES
LATCHING NO DEFEAT NO
1
2
3
4
5
ESC
UP
SELEC
DOW N
ACCPT
VALVES: ANALYZE
T=25C
C =9:59
A LA R M S SETU P
SELEC T A LA R M TO SETU P
1
2
3
4
5
A L 2
A L 1
M O R E
ESC
VALVES: ANALYZE
T=25C
C=9:59
FLOW = 150.0 CC/M IN
AUTORANGE: 10 PPM
1
2
3
4
5
RANGE
ALRM R
5.00
PPM 02
ALRM 1-HI
SETUP
STDBY
CAL