Zero calibration, Quic – Teledyne 3060e - Ultra Trace oxygen analyzer User Manual
Page 133
Teledyne Electronic Technologies
Analytical Instruments
6-9
Quic
Quic
Quic
Quic
Quick Ref
k Ref
k Ref
k Ref
k Reference
erence
erence
erence
erence
Ultra
Ultra
Ultra
Ultra
Ultra T
T
T
T
Trace Oxyg
race Oxyg
race Oxyg
race Oxyg
race Oxygen Anal
en Anal
en Anal
en Anal
en Analyz
yz
yz
yz
yzer
er
er
er
er
Figure 6-6. Calibration/Zero Calibration
CAL STDBY SETUP RANGE ALRMR
CAL
ZERO
THE SCREEN DISPLAYS WHETHER
THE ANALYZER IS USING THE
INTERNAL SCRUBBER
OR AN EXTERNAL ZERO GAS
PRESS ACK WHEN THE BACKGROUND
SIGNAL STABILIZES AND IS WITHIN
THE APPLICABLE LIMIT
SCREEN WILL DISPLAY PRESENT
OXYGEN CONCENTRATION, FLOW RATE
AND TIME REMAINING
MAN
CALIBRATION PROCESS COMPLETED
PRESS ACK TO RETURN TO ANALYZE MODE
PRESSING ANY KEY WILL ABORT
AUTO SETTLING
PRESSING ACK AFTER
ABORTING AUTO SETTLING
RETURNS YOU TO ANALYZE MODE
If the background level does not
fall below the acceptable limit after
5-7 days, replace the electrolyte with
prepurged electrolyte. Refer to Sections
3.3.1 and 4.3.7.2 for more details
NOTE:
AFTER AUTO SETTLING, THE SYSTEM
WILL AUTOMATICALLY RETURN
TO THE ANALYZE MODE
NOTE: Pressing ESC brings you
back to previous screen.
Zero Calibration