Model 3060e, Ultra trace sensor electrical connections – Teledyne 3060e - Ultra Trace oxygen analyzer User Manual
Page 34

3-8
Model 3060E
Model 3060E
Model 3060E
Model 3060E
Model 3060E
Chapter 3
Chapter 3
Chapter 3
Chapter 3
Chapter 3
Teledyne Electronic Technologies
Analytical Instruments
Figure 3-5. Inserting the Anode
Ultra Trace Sensor Electrical Connections
There are two wires to be connected so that the sensor can send signals to the
central processor to generate the O
2
content data, and two wires to send
temperature data to the processor.
The cathode terminal is located at the bottom of the sensor body and connected
to the red wire. The anode connection is on top of the sensor and is connected
to the black wire. The thermistor terminal is mounted on the stainless steel plate
on the top of the sensor body. With the exception of the anode, each of these
terminals is already connected to the signal processing unit when the analyzer
is shipped.
The positive and negative current source leads for the calibrator are located on
the top of the calibrator assembly (Figure 3-6).
After installing the analyzer, examine all of the connectors to ensure that they
are firmly secured to their respective terminals.
TOP
ANODE
PORT
ELECTROLYTE
CELL BODY