Troubleshooting Tables
Problem
Problem
Problem
Problem
Problem
Table
Table
Table
Table
Table
No Signal to Display/No Analog Output
5-1
Low Signal to Noise Ratio and Spikes on Signal
5-2
Background Signal > Maximum Correctable Value 5-3
Slow Response/Recovery Time
5-4
Temperature Failure < 5°C or > 45°C
5-5
Flow Failure Less than 20 cc
5-6
Calibration Current Failure
5-7
High Current Failure >30 PPM
5-8
Sensor Too Weak Error
5-9
Sensor Too Strong Error
5-10
No Span Gas From Int. Calibrator
5-11
NO SIGNAL
TO DISPLAY
CHECK SENSOR
ELECTRICAL
CONNECTIONS
CHECK ELECTROLYTE
LEVELS IN ULTRA
TRACE SENSOR
1
2
PROCEED
YES
FAILURE
CORRECTED
NO
NO ANALOG
OUTPUT SIGNAL
CHECK
CABLE
CONNECTIONS
FROM REAR OF CONTROL
MODULE TO OUTPUT
TERMINALS
CALL
SERVICE
Table 5-1. No Signal to Display/No Analog Output Signal