Teledyne Electronic Technologies
Analytical Instruments
Before zeroing background
signal = 60 ppb
After zeroing, background
signal = 0.0 ppb
Immediately after entering
Zero mode signal = 65 ppb
New background signal
Oxygen in
Analyze mode = 5ppb
65
60
5
Figure 4-8. Zero Mode displays a signal that is the
sum of the previously corrected background
plus the current oxygen level.
Setup Mode: System Password
Before you can access any of the system setup functions such as alarms, auto
settling, zero and span scheduling, or modem, and you do not want anyone else
to change any of the functions set, you must enter the authorization code. This
must be done before setting up any of the functions.
CAUTION:
Every time a COLD START is used, the code is automatically
erased from memory and must be reset; a WARM START will
not erase the code; nor will it reset the code.
1.
From the Analyze mode menu,
select
SETUP
.
VALVES: ANALYZE
FLO W = 150.0 CC/M IN
AUTORANGE: 10 PPM
1
2
3
4
5
RANGE
ALRM R
5.00
PPM 02
T=25C
C=9:59
SETUP
STDBY
CAL