Installation – Teledyne 3060e - Ultra Trace oxygen analyzer User Manual
Page 39

Teledyne Electronic Technologies
Analytical Instruments
3-13
Installation
Installation
Installation
Installation
Installation
Ultra
Ultra
Ultra
Ultra
Ultra T
T
T
T
Trace Oxyg
race Oxyg
race Oxyg
race Oxyg
race Oxygen Anal
en Anal
en Anal
en Anal
en Analyz
yz
yz
yz
yzer
er
er
er
er
Figure 3-10. Installing the Calibrator.
NOTE: DO NOT wipe the wick with a tissue paper.
8. Place the calibrator electrode assembly back on the stainless steel
block.
9. Using the wrench (P/N W68) provided, secure the calibrator elec-
trode assembly on the stainless steel block by tightening the four
bolts and then tighten the bolts with the wrench (P/N W69) pro-
vided. See Figure 3-10.
10. Reconnect the current source connector to the calibrator.
PPB OXYGEN
POWER
ON
CONTRO
L MODUL
E
TELEDY
NE
ANALYTIC
AL INSTR
UMENTS
PPM OXY
GEN
Calibrator
current source
connector
Figure 3-9. Soaking the Calibrator Wick with the Electrolyte.