Calibration – Teledyne LXT-280 User Manual

Page 29

Advertising
background image

Dissolved Oxygen Transmitter

Calibration

Teledyne Analytical Instruments

19

Calibration

4.1 Calibration

A complete calibration of the LXT-280 Dissolved Oxygen

transmitter involves:

1. Zero calibration
2. Span calibration
3. Output calibration
4. DO Locus calibration
5. Temperature calibration (trim)

In general, once the instrument has been setup, only step 2 will be

required for calibration although it is recommended that the output and
temperature calibration be occasional rechecked and adjusted as
necessary. It is also highly recommended that you perform a zero
calibration (step 1) periodically to check for drift and whenever a new
sensor is installed.

It may be useful to set the contrast of the display prior to calibration

to compensate for current lighting conditions.

Note: Before calibrating the LXT-280, place the instrument in

Manual Mode. See the following section.

4.2 Manual Output Mode

In order to perform calibrations without interfering with control or

recorder functions, the LXT-280 incorporates a manual output mode. In
the manual output mode, the current output is set to the desired level and
saved until changed or released from the manual mode. The following
procedure demonstrates the use of the manual output.
Procedure:

1. If not at the Main Menu, magnetically activate the appropriate

VERTICAL ▲/▼ arrow key.

Advertising