Dissolved Oxygen Transmitter

Calibration

Teledyne Analytical Instruments

25

4.8 Back-to-Factory Calibration

This feature allows the LXT-280 to be reset to "nominal electrode"

situation. This condition is an electrode with zero millivolts offset at
zero percent or ppm oxygen, and a full scale output of –400 mV. This
factory calibration feature is very useful in providing a reliable starting
point if the instrument has been mis-calibrated.

CAUTION: IDEAL INSTRUMENTATION CALIBRATION DOES

NOT IMPLY IDEAL SYSTEM CALIBRATION.


Before using the factory calibration feature, it must be understood

that it provides a nominal instrument calibration and does not allow for
electrode variations or shortened span of the electrode in service. In
order to obtain ideal system calibration, a span calibration is required.

The following procedure for returning the LXT-280 back-to-factory

calibration can be used for both zero and span calibrations.

This procedure is required at BOTH of the calibration points.

Procedure at the Zero Menu:

6. Magnetically activate the appropriate VERTICAL ▲/▼ arrow

key to reach the Electrode Zero Menu.

7. Position the cursor under the “C” in “Cal.”
8. Magnetically activate the RIGHT HORIZONTAL ►arrow key

and observe the cursor briefly move to the right and the mV
value change to 0.00. The default setting is complete.

9. Verify the calibration by observing a .0 calibration point at .0

mV.




Procedure at the Span Menu:

1. Magnetically activate the appropriate VERTICAL ▲/▼ arrow

key to reach the Electrode Span Menu.

2. Position the cursor under the “C” in “Cal.”

2 DO

.0

Cal

-400.0 mV