Dissolved Oxygen Transmitter

Calibration Correction

Teledyne Analytical Instruments

33

Table 5-2: Relative Humidity Correction Factors

Temperature in °C

Relative

Humidity

0

5

10

15

20

25

30

35

0%

1.000 1.000 1.000 1.000 1.000 1.000 1.000 1.000

10%

.999

.999

.999

.998

.998

.997

.996

.994

20%

.999

.998

.998

.997

.995

.994

.992

.989

30%

.998

.997

.996

.995

.993

.991

.987

.983

40%

.998

.997

.995

.993

.991

.987

.983

.978

50%

.997

.996

.994

.992

.988

.984

.979

.972

60%

.996

.995

.993

.990

.986

.981

.975

.967

70%

.996

.994

.992

.988

.984

.978

.971

.961

80%

.995

.993

.990

.987

.982

.975

.967

.956

90%

.995

.992

.989

.985

.979

.972

.962

.950

100%

.994

.991

.988

.983

.977

.969

.958

.945

Table 5-3: Elevation Correction Factor

Elevation Correction Factor

Feet

0

1,000 2,000 3,000 4,000 5,000 6,000 7,000

0

1.000

.964

.930

.897

.865

.834

.804

.775

100

.996

.961

.926

.893

.861

.831

.801

.772

200

.993

.957

.923

.890

.858

.828

.798

.770

300

.989

.954

.920

.887

.855

.825

.795

.767

400

.986

.950

.916

.884

.852

.822

.792

.764

500

.982

.947

.913

. 880

.849

.819

.788

.761

600

.978

.943

.910

.877

.846

.816

.787

.758

700

.975

.940

.906

.874

.843

.813

.784

.756

800

.971

.937

.903

.871

.840

.810

.781

.753

900

.968

.933

.900

.868

.837

.807

.778 .750

Advertising