Calibration correction factors – Teledyne LXT-280 User Manual

Page 41

Advertising
background image

Dissolved Oxygen Transmitter

Calibration Correction

Teledyne Analytical Instruments

31

Calibration Correction Factors

5.1 Calibration Correction Factors

Temperature, relative humidity, elevation and salinity factors affect

the solubility of oxygen in a liquid and thus affect the accuracy of the
dissolved oxygen calibration using the air calibration method. The affect
on solubility is caused by barometric pressure changes due to changes in
the atmospheric environment. Because the dissolved oxygen sensor
responds to pressure changes, the output of the sensor will be affected
by these pressure variations. To correct for these factors, refer to the
tables below.

Please note that the tables only apply to absolute dissolved oxygen

measurements (those in ppm or mg/l and not to relative dissolved
oxygen measurements (typically expressed as percent oxygen). The data
is used when performing an air span calibration or an air saturated
sample calibration.

To compensate for these effects, use the following procedure:

1. Determine the temperature of the air (if the electrode is being

calibrated in air only) or the temperature of the air saturated
sample (if the electrode is being calibrated in solution).

2. Refer to Table 5-1, Oxygen Saturation vs. Temperature and find

the solubility value of oxygen in ppm for that temperature.
Interpolation may be necessary if the temperature is between the
listed values. Record this value.


Advertising