Teledyne LXT-280 User Manual

Page 33

Advertising
background image

Dissolved Oxygen Transmitter

Calibration

Teledyne Analytical Instruments

23

2. Magnetically activate the DOWN VERTICAL ▼ arrow key to

reach the Electrode Span Menu and locate the cursor under the
"2".

3. For diagnostic reasons note the slope values from the prior

calibration.

4. Magnetically activate the RIGHT HORIZONTAL ►arrow key

to enter the calibrate mode. Observe the diagnostic value on the
bottom line change from historical data to a real-time value and
that the term "Cal" has disappeared.

5. To change the buffer point, position the cursor under the desired

digit using the LEFT HORIZONTAL◄ arrow key.

6. Increase or decrease the value by magnetically activating the

appropriate VERTICAL ▲/▼ arrow key.

7. Wait for the real-time mV value to stabilize.
8. To file the calibration and exit the calibrate mode, magnetically

activate the LEFT HORIZONTAL◄ arrow key until the cursor
is all the way to the left and the term “Cal” on the second line
reappears.

A typical DO electrode will indicate a full scale value of –400 mV.

See the equation in Section 3.2.3 for information on determining the
effective range of the electrode.

CAUTION: IF A FLASHING ASTERISK APPEARS DURING OR

AFTER CALIBRATION, THE ELECTRODE MAY
REQUIRE SERVICE, OR THE BUFFER SOLUTION
MAY BE CONTAMINATED. SEE SECTION 4.7.

Comparing the millivolt values before and after the calibration

provides a valuable diagnostic tool in determining the degradation of the
electrode. Minor differences in the full scale value is normal; however,

2 DO

100.0

mv/dec