Teledyne LXT-280 User Manual

Page 36

Advertising
background image

Calibration

LXT-280

Teledyne Analytical Instruments

26

3. Magnetically activate the RIGHT HORIZONTAL ►arrow key

and observe the cursor briefly move to the right and the full scale
millivolt value changes to –400.0. The default setting is
complete.

4. Verify the calibration by observing a .0 calibration point at –400

mV/FS.



4.9 Output Calibration

The Output Calibration screen is used to define the 4-20 mA range

for the measurement. The values in the menu are in the units being
measured. Maximum ranges are 0-200 % and 0-20 ppm.

To change or expand the 4-20 mA range, use the following

procedure.
Procedure:

1. Magnetically activate the appropriate VERTICAL▲/▼ arrow

key to reach the Output Calibration Menu.

2. Position the cursor on the top line to change the 4 mA point or

the bottom line to change the 20 mA point.

3. Magnetically activate the RIGHT HORIZONTAL ► arrow key

to enter the calibrate mode.

4. To change the value, position the cursor under the desired digit

using the left LEFT HORIZONTAL◄ arrow key.

5. Increase or decrease the value by magnetically activating the

appropriate VERTICAL ▲/▼ arrow key.

6. When the desired value is reached, save the calibration and exit

the calibrate mode by magnetically activating the LEFT
HORIZONTAL ◄ arrow key until the cursor is all the way to
the left.

2 DO

.0

Cal

-400.0

Advertising