Introduction

The Titan scale was designed to be used with Intalogix™ systems. Intalogix™ systems
utilize smart sectional controllers (SSC) and pit power supplies (PPS) for load cell
excitation and signal processing.

Analog instruments cannot be used with this platform. The sensitivity using a analog
indicator would be approximately a half (.5) microvolt. Most analog instruments have a
minimum sensitivity of one (1) microvolt.

B. Description

There is one (1) SSC per section and one (1) PPS for the entire platform unless the
number and resistance of the cells require a second pit power supply. Smart sectional
controller boxes have four (4) terminals: two (2) for load cells and two (2) for interfacing
to other SSC boxes or terminating to a pit power supply. All cell/section/scale
adjustments are made via the Intalogix™ system instrument.

C. Installation

a.

Boxes

The box has mounting brackets which allow mounting to the side of the Titan
modules.

b.

Wiring

Cable used in all wiring must be a minimum of 18 AWG. Use cable 17204 or 17246.
Use appropriate service manual for the indicator being installed or refer to the
Appendix for typical wiring information.

50745

14

03/10 Rev. 5

Section 4: Electrical Installation