But, now we want to look at fall times greater than 45.5 ns; 3 falling edges
from the same acquisition pass this new filter:
You can also decide how anomalous the features located should be by
using the
to select the average value as the limit, to search
for rarest events, or to apply statistical criteria.
F
ILTER
W
IZARD
You can decide how anomalous the features searched or scanned for
should be by using the average value as the limit, by searching for rarest
events, or by applying statistical criteria: ±1, 3, or 5 sigma.
In the following example, there were several slow edges in a repetitive
waveform. For each filtering method selectable from the wizard, the rise
time used as the filter limit and delta are calculated automatically.