Remote control commands and queries, Cursor – Teledyne LeCroy X-STREAM OSCILLOSCOPES Remote Control User Manual
Page 215

Remote Control Commands and Queries
WM-RCM-E Rev D
ISSUED: February 2005
209
CURSOR
PER_CURSOR_SET, PECS
Command/Query
DESCRIPTION
The PER_CURSOR_SET command allows you to position one or
more of the six independent cursors at a given grid location. Cursors
must be turned on for the PECS? query to work.
NOTATION
HABS
Horizontal absolute
VABS
Vertical absolute
HDIF
Horizontal difference
VDIF
Vertical difference
HREF
Horizontal reference
VREF
Vertical reference
COMMAND SYNTAX
<trace>:PEr_Cursor_Set
<cursor>,<position>[,<cursor>,<position>,…,<cursor>,<position
>]
<trace>:={C1, C2, C3, C4, F1 through F8, M1, M2, M3, M4, TA,
TB, TC, TD}. TA through TD are included for compatibility with
existing software for earlier instruments. These four mnemonics will
not be used in responses to queries.
QUERY SYNTAX
<trace>:PEr_Cursor_Set?
EXAMPLE(GPIB)
The following instruction positions the HREF and HDIF cursors at
+2.6 DIV and +7.4 DIV respectively, using Channel 2 as a reference:
CMD$=”C2:PECS HREF,2.6 DIV,HDIF,7.4 DIV”
CALL IBWRT(SCOPE%,CMD$)
RELATED COMMANDS
CURSOR_MEASURE, CURSOR_SET, PER_CURSOR_VALUE