Remote control commands and queries, Acquisition – Teledyne LeCroy X-STREAM OSCILLOSCOPES Remote Control User Manual
Page 99

Remote Control Commands and Queries
WM-RCM-E Rev D
ISSUED: February 2005
93
ACQUISITION
ATTENUATION, ATTN
Command/Query
DESCRIPTION
The ATTENUATION command selects the vertical attenuation factor
of the probe. Values up to 10000 can be specified.
The ATTENUATION? query returns the attenuation factor of the
specified channel.
COMMAND SYNTAX
<channel> : ATTeNuation <attenuation>
<channel> : = {C1, C2, C3, C4, EX, EX10}
<attenuation> : = {1, 2, 5, 10, 20, 25, 50, 100, 200,
500
, 1000, 10000}
QUERY SYNTAX
<channel> : ATTeNuation?
RESPONSE FORMAT
<channel> : ATTeNuation <attenuation>
AVAILABILITY
<channel> : {C3, C4} available only on four-channel oscilloscopes.
EXAMPLE (GPIB)
The following instruction sets to 100 the attenuation factor of
Channel 1:
CMD$=“C1:ATTN 100”: CALL IBWRT(SCOPE%,CMD$)