Troubleshooting
178
Active Session Conflict
Two active software sessions when attached to the same parallel port will cause the
TA700/800 to operate improperly.
Recommendations:
1. Make sure that there is only one active session.
2. For applications that require two TA700/800 analyzers and two software
sessions running in the same system, you may implement that by using two
parallel ports with each session connected to its own port.
To implement this:
1. Run the first session with only one TA700/800 connected to a parallel port.
2. Disconnect the TA700/800 from the parallel port.
3. Connect the second TA700/800 to the second parallel port and start another
software session.
4. Reconnect the first TA700/800.
System Clock Shutdown
Some host systems may shut down the system clock to the TA700/800 slot if the
TA700/800 is set to be transparent to the BIOS.
Known Issues
For the most current known issues and problem resolutions visit
http://www.getcatalyst.com