Reading Configuration (If Process was selected)

INPUT/READING (SCALE AND OFFSET) SUBMENU:

Input Voltage or Current can be converted or scaled into values appropriate for
the process or signal being measured. So, a reading may be displayed, for
example, in units of weight or velocity instead of in amperes or volts.

The instrument determines Scale and Offset values based on two user-provided
input values entered with the corresponding readings. Note that “In1” Input 1 and
“In2” Input 2 are represented and entered as a product of the input
voltage/current and the conversion number from the Table 3.1.

The following instructions include details for a specific scenario in which a
4-20 mA input (in the 20 mA Process Mode) is to be represented as a
measurement of 0-100 percent.

Press

d

14) Press

d at the IN.RD prompt. Display shows IN1 Input 1

Submenu.

Press

d

15) Display shows Input 1 value with 1

st

digit flashing.

Press

b & c 16) Use b and c buttons to enter IN1 value.

The

IN1 value = min. input value * conversion number.

Disregard the position of the decimal point (2000 counts may
actually appear as “2000”, “200.0”, “20.00”, or “2.000”).

Example: 4 mA as 4(mA) x 500 = 2000.

Press

d

17) Display advances to

RD1 Reading 1 Submenu.

Press

b & c 18) Use b and c buttons to enter RD1 value.

This value represents

IN1 in terms of some meaningful

engineering units. To show the 4 mA as zero percent enter

RD1

value = 0000.
Example:

RD1 value = 0000.

Press

d

19) Display

IN2 Input 2 Submenu.

Press

d

20) Display shows Input 2 value with 1

st

digit flashing.

The

IN2 value = max. input value * conversion number.

Example: 20(mA) x 500 = 10000 (9999).

Press

b & c 21) Use b and c buttons to enter IN2 value.

Press

d

22) Display advances to

RD2 Reading 2 Submenu.

Press

b & c 23) Use b and c buttons to enter RD2 value.

Example:

RD2 value = 0100.

Press

d

24) Display flashes

STRD stored message momentarily and

then advances to

ALR1 only, if change was made, otherwise

press

d to advance to ALR1 Alarm 1 Menu.