NEWPORT iDRA Controller User Manual

Page 28

Advertising
background image

24

Conversion number is a coefficient of conversion between input values
and real full display range (10000 counts, shown as 9999). See Table 3.2
below for proper conversion number.

Table 3.2 Conversion Table

RANGE

CONVERSION NUMBER

100 mV

10000 / (100 x 1) = 100

1 V

10000 / (1000 x 1) = 10

10 V

10000 / (1000 x 10) = 1

0 -20 mA

10000 / (20 x 1) = 500

Example =

0 - 1 V = 0 - 100.0

In 1 = 0

Rd 1 = 0

Inp 2 = 9999

Rd 2 = 100.0

Advertising