E Series - 12 Lead ECG Monitoring Option Insert

16

9650-1213-01 Rev. G

Operationally, there is no difference in the way MUSE transmissions are handled with the satellite phone versus the
standard fax modem card and cell phone. You perform an Acquire or Select Patient Record operation, select a phone
number, and press the Transmit softkey. However, you must understand the characteristics of satellite phone
behavior. Transmissions must have an unobstructed line of sight to a satellite. This means that indoor transmissions
will most likely not work. The satellite phone should generally be used outdoors or have an outdoor antenna.

Note that the display on the satellite phone should be used as an indicator of dialing progress. If a dialing sequence
stalls, you should abort the dialing sequence and manually restart the transmission on the E Series unit.

When you attempt to fax data via satellite phone, the E Series detects the presence of the RS232 card and issues a
general fax error.

Finally, you should also note that data card uploads are not handled via the National Instruments PCMCIA RS232
card. This function is accessed via the RS232 port on the back of the E Series unit. Dial-up clock synchronization is not
available when the National Instruments PCMCIA RS232 card is installed.

Transmission Specifications

Fax

Group 3 Facsimile

FAX Software Interfaces

Class 2.0

Class 2

Class 1

MUSE

MUSE revision 4B and higher

Telephone Company Requirements

The following regulations apply to modems used within the United States and Canada.

FCC Regulations

The FCC has established rules that permit this device to be directly connected to the telephone network, using a
standardized jack. Do not use this equipment on a party line or coin line.

Malfunctioning equipment may cause damage to the telephone network. If this device is not functioning properly,
disconnect it until the problem has been determined and the device has been repaired. Otherwise, the telephone
company may disconnect service temporarily.

The modem card is a non-serviceable/repairable item. It is your responsibility to report the need for any service to
the device to ZOLL Medical Corporation.

If you encounter any problems with your telephone after installing any new device, disconnect it from the telephone
line to see if the device is the source of the problem.

The telephone company may make changes to its technical operations and procedures. If such changes affect the
compatibility or use of this device, the telephone company is required to provide adequate notice of the changes.

Phone Company Requests

If the telephone company requests information about the equipment connected to their lines, inform them of:

CAUTION

Data transmission via cellular phones can be less reliable than transmission via landline connections. A strong
signal and stationary transmission will improve the transmission’s success rate. Follow the directions provided with
your cellular phone.

Many hospitals prohibit the use of cellular phones on their premises. Please abide by local rules and regulations.