12-Lead Acquisition

9650-1213-01 Rev. G

5

mode; see the E Series Configuration Guide for more
information.)

Once the settings screen is displayed, press the
Settings softkey again to scroll through and highlight the
different available selections. Press the Enter softkey to
select the setting that is highlighted. You can return to
the 12-lead Monitor screen by pressing the Return
softkey.

Note: AED units in Semi-Automatic mode do not

display “Filter” or “Lead Grp.” Only “Dial Type”
displays.

Filter Setting

Selecting Filter allows you to select among the following
three filter/print format settings for 12-lead monitoring:

0.05 - 150 Hz frequency 4x3 channel printout:
With this (0.05 - 150Hz 4x3) setting, all 4x3 12-lead
ECG reports print with a 0.05 to 150 Hz bandwidth.

0.05 - 150 Hz frequency continuous printout:
With this (0.05 - 150Hz Cont.) setting, pressing and
holding the RECORDER button prints a continuous
ECG strip using full diagnostic bandwidth until the
RECORDER button is released.

Note: Use a diagnostic bandwidth of 0.05 to 150 Hz for

diagnostic ECGs. ZOLL Medical Corporation
suggests using a 0.05 to 40 Hz setting in
environments prone to signal noise.

0.05 - 40 4x3 channel printout (0.05 - 40Hz 4x3):
With this setting, all 4x3 12-Lead ECG reports print
with a 0.05 to 40 Hz bandwidth. Use this setting to
reduce excessive muscle or other artifact. This
setting does not alter the bandwidth of ECG data
passed to the 12SL program; it only affects the
printed waveforms.

Select the filter you want to use by pressing the Filter
softkey. The highlight area scrolls among the different
filter choices. Press the Enter softkey to save the
highlighted filter and return to the 12 Lead Monitor sub-
menu.

Note: When the E Series is turned off for more than 10

seconds, all settings are restored to their default
settings.

Lead Group Setting

The E Series 12-Lead option allows any three ECG lead
signals to be simultaneously printed on the stripchart
recorder when using a 12-lead cable. You can choose
options Standard, Custom 1 or Custom 2, as follows:

Standard

When programmed to this setting, 3-lead ECG
recordings are grouped and printed together as
follows:

— leads I, II, and III

— leads aVR, aVL, and aVF

— leads V1, V2, and V3

— leads V4, V5, and V6

The lead group that prints on the recorder in this
mode is the group containing the lead selected for
display on the E Series screen. For example, if lead II
is displayed, leads I, II and III are printed; if lead V3 is
displayed, leads V1, V2, and V3 are printed.

Note: When the E Series is turned off for more than 10

seconds, all settings are restored to their default
values.

Custom

When the lead group is configured to either Custom 1
or Custom 2, 3-lead ECG recordings include the
leads preconfigured for that particular custom group.
(See the E Series Configuration Guide for more
information.)

Press the Lead Grp softkey to scroll through and
highlight the different Lead Group selections for three
channel printing.

Press the Enter softkey to save the highlighted Lead
Group and return to the 12 Lead Monitor menu.

Dial Type Setting

Selecting the Dial Type setting allows you to select
either Tone or Pulse dialing.