Energy level: shock 1, 2, 3, Auto analyze 3 times, No shock advised" prompt – ZOLL M Series CCT Defibrillator Rev A User Manual

Page 16: Check patient" prompt, Display do cpr message, Display ecg in semi-auto mode, Nergy, Evel, Hock, Nalyze

Advertising
background image

M S

ERIES

CCT C

ONFIGURATION

G

UIDE

12

Energy Level: Shock 1, 2, 3

Selects the energy levels used for the 1

st

, 2

nd

, and 3

rd

consecutive shocks that the defibrillator

delivers when:

In semi-automatic or advisory mode and using pads

Ιn manual mode, using external paddles or pads, with Basic Auto Energy Escalation set
to “Yes”

Energy level 1 is the energy level the unit defaults to when the power is turned on with external

paddles or pads connected. For more information about the Basic Auto Energy Escalation

setting, see the “Basic Auto Energy Escalation” section of this manual.

Auto Analyze 3 Times

Sets the unit to automatically repeat ECG rhythm analyses after the delivery of shocks in the

AED and Advisory Modes. When set to “Yes”, the unit automatically analyzes the patient’s

ECG rhythm after the first and second shocks in each three-shock sequence. Following delivery

of the third shock automatic reanalysis is halted and the “CHECK PULSE” message is

displayed. This indicates that the three shocks sequence has been completed and the patient's

pulse should be checked. This function only operates when the “Auto Charge in Advisory

Mode” configuration option is also enabled.
When set to “Only in Semi-auto,” the unit functions as described above only when it is in AED

semi-auto mode. If the unit is changed to Manual mode when set to “Only in Semi-auto”, any

automatic analysis/reanalysis sequencing terminates. If the Analyze button is pressed, the unit

analyzes only once.

"NO SHOCK ADVISED" Prompt

Changes the display message and the voice prompt issued after analysis of non-shockable

rhythm.

"CHECK PATIENT" Prompt

Changes the display message and the voice prompt issued when a shockable rhythm has been

detected by background ECG analysis and Heart Rate Alarms are enabled, or the unit is

operating in Semi-Automatic mode.

Display DO CPR Message

Sets unit to display messages “IF NO PULSE” / “PERFORM CPR” alternately for one minute

after each set of stacked shocks even if a perfusing rhythm is present.
ECG Rhythm analysis will restart automatically 70 seconds after delivery of the third shock in a

three-analysis/shock sequence. The “CHECK PATIENT” message and voice prompt will also

be inhibited for 70 seconds, following completion of a user activated analysis or discharge. This

feature is operational only when the “Auto Charge in Advisory Mode” and “Auto Analyze 3

times” configuration options are enabled.
When set to “Only in Semi-Auto,” the unit functions as described above only when it is in AED

semi-auto mode.

Display ECG in Semi-Auto Mode

Sets the device to display or not display the patient’s ECG trace on the monitor while in

semiautomatic mode.