Configuration Settings

13

Store to Card in Auto Monitor Mode (AED models)

Sets unit to automatically store patient’s ECG trace and voice data to the PCMCIA card when:

ECG

leads are attached to the patient

MFE pads are not connected

M Seri

es CCT unit is on

If disabled, the M Series CCT unit will not record any data to the PCMCIA card unless MFE

pads are connected to the patient.

Auto Charge in Advisory Mode

When enabled and a shockable rhythm is detected after pressing the ANALYZE button, the

unit will automatically charge the defibrillator to the pre-configured or user selected energy

setting.

Record to Card in Advisory Mode

Sets the unit to record ECG data (ECG only), ECG and voice data (Voice and ECG), or nothing

(None) to the memory card.

Enable Voice In Advisory Mode

Enables or disables voice prompts when operating in Advisory or Manual operating modes. If

enabled, the voice prompts will sound at the appropriate time when in Advisory mode or

Manual mode.

Display Heart Rate in Semi-Auto Mode

Sets the device to display or not display the patient heart rate on the monitor while in

semiautomatic mode.

Display Lead on Power-Up

Allows the user to select any one of the following Leads to be displayed on power-up: Paddles/

Pads, I, II, or III. When the unit powers up in semi-automatic mode, the option is ignored. The

unit will power up in Lead II, and switch leads automatically between leads II and MFE for

Auto Monitor and Auto Defib modes, respectively. When the unit enters Manual Mode (Manual

Mode override sequence is completed), the lead switches to the setting specified in the "Display

Lead on Power Up" option.

Frequency Response

Selects the ECG filter bandwidth to be used during ECG monitoring. The frequency of the

device can be set to either help reduce noise, help with the diagnostic response, or set as a

default response.

Enable 12-Lead in Semi-Auto Mode (AED with 12-Lead)

If set to “Yes,” 12-lead analyses, transmissions and print-outs are available in semi-automatic

(AED) Auto Monitor mode (ECG leads on the patient with MFE Pads disconnected), depending

on other configuration option settings. If set to “No,” 12-lead functions are available in Manual

Mode only.

QRS Volume on Power-Up

Turns the default QRS beeper volume to midrange or off following power up of the unit.
If set to “Off in Semi-Auto Mode,” the QRS beeper volume turns off in semi-auto mode. If the

unit is changed to manual mode, the beeper volume sets to midrange.