1

Introduction

The M Series CCT is designed with several user selectable features that allow operators to

configure the device according to their protocols and local requirements. This manual describes

the configuration options and how to change them. This manual also lists the configurable

features, options and default settings. Some features are available to manual mode operators

only.
The default settings shown below are for North American units only. To print the current

configuration of your unit, while the unit is in the configuration mode, press the Print Config

softkey.

Configuring the Unit

After the unit has been off for 10 seconds or more, simultaneously press and hold the rightmost

and leftmost softkeys. Turn the selector switch to MONITOR or ON while holding the buttons

depressed for approximately four (4) seconds. After four (4) seconds the following screen will

appear.

Users entering System Configuration for the first time must enter the default passcode

(00000000) for System Configuration to continue.
Press the 0 (zero) softkey. This moves the highlight to the next digit. Repeatedly press the 0

(zero) softkey until all eight digits have been entered. The Configuration Passcode screen will

then automatically advance to the Select Language menu.
To change one of the entered digits use the Delete softkey to highlight the incorrect digit and

reenter the correct code.

CAUTION

The following features are not currently implemented on the M Series CCT: Voice Markers and

Auto Self-Test.