Connecting the hose, Applying the cuff to the patient, Accessing nibp features – ZOLL M Series Defibrillator Rev G NIBP User Manual

Page 12

Advertising
background image

N

ON

-I

NVASIVE

B

LOOD

P

RESSURE

(NIBP)

6

ZOLL M Series Operator’s Guide - Option Insert

Connecting
the Hose

The NIBP option comes with a hose that has a plastic connector on one end and a metal connector on
the other end. Attach this hose to both the M Series unit and the cuff before taking a measurement. The
cuff has its own short length of hose with a connector on the end. This connector fits into the end of
the hose that is not connected to the M Series unit.
To connect the hose:
1. Insert the metal connector into the NIBP connector at the front of the M Series unit and push the

connector until it snaps into place.

2. Insert the plastic connector into the cuff connector and twist the connectors until they lock into

place.

You can now apply the cuff to the patient.

Applying the
Cuff to the
Patient

Using a cuff that is too small or is loosely applied results in values higher than the patient’s actual
blood pressure. Using a cuff that is too large results in values lower than the patient’s actual blood
pressure.
To apply the cuff to the patient:
1. Ensure the patient is sitting or lying down with the limb relaxed, extended and placed on a smooth

surface for support.

2. Place the cuff 2 to 5 cm above the elbow crease or 5 to 10 cm above the knee crease.

CAUTION

Do not place the NIBP cuff on the same arm/leg as an SpO

2

sensor. Inflation of the cuff will cause the

SpO

2

monitor to read incorrectly.

CAUTION

Ensure that the cuff is at the same level as the heart. If the cuff is located below the patient’s heart, the

blood pressure measurements display false high readings. If the cuff is located above the patient’s
heart, the blood pressure measurements display false low readings.

3. Adjust the cuff so that the artery marker on the cuff is over the artery, pointing to the hand or foot.
4. Check that the cuff ends between the Range lines marked on the cuff. If they do not line up, use a

different sized cuff.

5. Wrap the deflated cuff snugly around the limb without impeding blood flow.
6. Ensure that the hose is routed to avoid kinking or compression.
You can now access the NIBP features.

Accessing NIBP Features

Unless you are sure that the NIBP cuff inflation and alarm settings are appropriate for the patient,
display the NIBP menu before you take a blood pressure measurement. When you first turn on the
M Series unit, the NIBP settings are at their default values.
While the factory-installed default settings are appropriate for most patients, do not assume the
settings are at their default. NIBP settings may not be at their factory-installed defaults because a
previous user may have:
• Changed the settings (if you did not turn on the M Series unit).
• Reconfigured the default settings.
Use the default settings unless they are clearly inappropriate for the patient. Any changes to these
settings remain in effect until either the settings are changed or 10 seconds after the M Series unit is
turned off. If you have not received training on setting NIBP features, do not use the NIBP option.
To facilitate quick reaction during emergency situations, you can directly access many NIBP features
without displaying the NIBP menu (see the following table).

Advertising