ZOLL M Series Defibrillator Rev G NIBP User Manual

Page 28

Advertising
background image

N

ON

-I

NVASIVE

B

LOOD

P

RESSURE

(NIBP)

22

ZOLL M Series Operator’s Guide - Option Insert

REPEAT NIBP MEAS
alternating with NIBP
SIGNAL WEAK

Weak or no oscillometric signal.

Check cuff fit and positioning.
Check hose connection to unit.
Check for kinked hose.
Increase cuff inflation pressure if clini-

cally appropriate.

SYSTEM FAULT

Communication problem with the NIBP

module.

Cycle power and retry.
If problem persists, return for service.

False high reading

Cuff too small.
Cuff not centered over brachial artery.
Cuff too loose.
Patient’s limb below level of heart.

Select larger cuff.
Readjust cuff.
Reapply cuff or select smaller cuff.
Raise patient’s limb to heart level.

False low reading

Cuff too large.
Limb above level of heart.

Select smaller cuff
Lower patient’s limb to heart level.

Message/Symptom

Problem

User Action