Turning off auto adjust, Nibp) – ZOLL M Series Defibrillator Rev G NIBP User Manual

Page 16

Advertising
background image

N

ON

-I

NVASIVE

B

LOOD

P

RESSURE

(NIBP)

10

ZOLL M Series Operator’s Guide - Option Insert

Figure 7 Cuff Inflation Menu

On M Series CCT units, the “Inflation Pressure” field is labeled “Adult Inflation Pressure”,
“Pediatric Inflation Pressure”, or “Neonatal Inflation Pressure” if the Patient Type is set to Adult,
Pediatric, or Neonatal respectively.

2. Press the Next softkey if “Inflation Pressure” is not highlighted.
3. Press the Inc or the Dec softkey to select cuff inflation pressure.
4. Press the Return softkey (twice on M Series CCT units) to return to the NIBP menu.

The cuff inflation pressure is set.

Turning Off
Auto Adjust

You should turn off the Auto Adjust feature in situations where the M Series unit might inflate the cuff
to levels that can cause discomfort or harm to the patient. For example:
• Patient is being transported over a very bumpy road.
• Patient is moving the monitored arm.
• Patient is a small child or infant.
When you turn the Auto Adjust feature off, the M Series unit repeats the same cuff inflation pressure
for every measurement. Because this method does not readjust to the patient’s previously measured
blood pressure, do not use fixed cuff inflation pressures unless you have received the proper training.

To turn off Auto Adjust:
1. Press the Cuff Inflation softkey from the NIBP menu. If the Cuff Inflation softkey is not

displayed, see "Displaying the NIBP Menu" on page 7.

NOTE

On M Series CCT units, you access the Cuff Inflation softkey from the NIBP Settings menu. Press
the Settings softkey from the NIBP menu, then press the Cuff Inflation softkey. If the Settings
softkey is not displayed, see "Displaying the NIBP Menu" on page 7.

The M Series unit displays the Cuff Inflation menu.
Figure 8 Cuff Inflation Menu

2. Press the Next softkey until “Auto Adjust” highlights.
3. Press the Inc softkey until “OFF” displays.

Figure 9 Auto Adjust Menu

4. Press the Return softkey (twice on M Series CCT units) to return to NIBP menu.

The M Series unit repeats the selected cuff inflation pressure for every measurement.

Inc

Dec

Prev

Next

Return

Inflation Pressure

Auto Adjust

160 mmHg

160 mmHg

On

j

g

Inc

Dec

Prev

Next

Return

Inflation Pressure

Auto Adjust

160 mmHg

160 mmHg

On

j

g

Inc

Dec

Prev

Next

Return

Inflation Pressure

Auto Adjust

160 mmHg

Off