Specifications

ZOLL M Series Operator’s Guide - Option Insert

23

Specifications

* Neonatal and pediatric NIBP modes available on M Series CCT units only.

a

Blood pressure measurements determined with this device are equivalent to those obtained by a trained observer using the

cuff/stethoscope auscultation method, within the limits prescribed by the American National Standard, Electronic or

automated sphygmomanometers (AAMI SP10). To receive a copy of the report containing the AAMI SP10 test results, contact

ZOLL Technical Support at (800) 348-9011 or (978) 421-9655

.

Table 8 General Specifications

Principle of Operation

Oscillometric

Warm up Time

Operational in less than 10 seconds

Size

8.2" (20.8 cm) high x 10.1" (25.7 cm) wide x 8.0" (20.3 cm) deep

Regulatory Standards

ANSI/AAMI SP10-1992;
EN 1060-1: 1996 Specification for Non-invasive sphygmomanometers;
EN 1060-3: 1997;
IEC60601-2-30: 1999

Table 9 NIBP Measurement Specifications

Range

Systolic

Adult: 40 – 260 mmHg; 5.3 – 34.7 kPa
Pediatric: 35 – 160 mmHg; 4.7 – 21.3 kPa*
Neonatal: 25 – 130 mmHg; 3.3 – 17.3 kPa*

Diastolic

Adult: 20 – 200 mmHg; 2.7 – 26.7 kPa
Pediatric: 20 – 130 mmHg; 2.7 – 17.3 kPa*
Neonatal: 20 – 105 mmHg; 2.7 – 14.0 kPa*

Mean

Adult: 20 – 220 mmHg; 2.7 – 29.3 kPa
Pediatric: 20 – 140 mmHg; 2.7 – 18.7 kPa*
Neonatal: 20 – 110 mmHg; 2.7 – 14.7 kPa*

Accuracy

Per AAMI SP10 +/- 5 mmHg (0.7 kPa) mean difference;
8 mmHg (1.1 kPa) standard deviation

a

Resolution

1 mmHg; 0.13 kPa

Measurement Cycle Time

Typical: 30 seconds

Worst Case: 180 seconds

Table 10 Pulse Rate Measurement Specifications

Range

35 – 220 pulses per minute

Resolution

1 pulse per minute