Troubleshooting – ZOLL M Series Defibrillator Rev G NIBP User Manual

Page 27

Advertising
background image

Troubleshooting

ZOLL M Series Operator’s Guide - Option Insert

21

Troubleshooting

The following table lists the NIBP messages, the problem each message indicates, and the associated
corrective action(s). Read this section carefully before monitoring patients. For further assistance, call
ZOLL Technical Service at 1-800-348-9011. International customers should call the nearest
authorized ZOLL Medical Corporation distributor.

Message/Symptom

Problem

User Action

CALIBRATE NIBP

NIBP calibration incomplete or failed

Recalibrate NIBP. (See M Series Service

Manual for details).

NIBP COMM ERROR

Communication problem with the NIBP

module.

Cycle power and retry.
If problem persists, return for service.

NIBP FAULT

No communication from the NIBP module.
NIBP module failed self-test.

Cycle power and retry.
If problem persists, return for service.

NIBP MEAS
ABORTED

Cuff inflation pressure set too high for

attached cuff.

Inflation too fast.
M Series unable to find systolic value for 180

seconds.

Defibrillator charged or charging. User-

initiated abort.

Confirm Defib was not charging.
Verify that you are using proper sized

cuff.

Check for cuff and hose blockages.
If problem persists, return for service.

NIBP NOT READY

Defib charged or charging in progress.
NIBP module performing self-test after

power-up.

Wait until Defib discharges or dumps

before taking the next measurement.

Wait for more than 10 seconds after

power up before taking blood
pressure measurements.

NIBP OUT OF
RANGE

The data from the NIBP module is out of

range.

Measure patient’s blood pressure with

other equipment.

Check cuff fit and positioning.
Switch cuff to other arm.
If problem persists, return for service.

REPEAT NIBP MEAS

Exceeded maximum number of inflation

attempts.

Exceeded 180-second measurement time

limit.

Check cuff and hose.
Repeat NIBP measurement.

REPEAT NIBP MEAS
alternating with
CHECK CUFF/HOSE

Blood pressure cuff or hose not installed

correctly.

Faulty cuff or hose.
Hose kinked or disconnected.
Inflation rate too fast or too slow.

Check cuff connection to hose.
Check hose connection to unit.
Check for kinked hose or air leaks in

hose.

If problem persists, replace cuff then

hose.

REPEAT NIBP MEAS
alternating with NIBP
ARTIFACT

Unable to detect systolic, diastolic or mean

blood pressure due to excessive motion or
vibration.

Take a single blood pressure

measurement.

Keep patient as still as possible.
Insulate patient and cuff from outside

vibrations as much as possible.