Lead reports, Lead ecg waveforms – ZOLL M Series Defibrillator Rev H User Manual

Page 10

Advertising
background image

OPTION INSERT

9650-0215-01

12-lead - 8

12-Lead Reports

Immediately after acquisition, the 12-lead data prints in the following order:

1. 12-Lead ECG waveforms

2. Patient

information

3. ECG analysis using the GE Medical Systems Information Technologies 12SL Analysis Program (if configured)

4. Measurements matrix (if configured)

See the figures below for a view of numbers 1, 2, and 4 above.

Optionally, the user can configure the unit to print median complexes in a 4x3 format or standard data in a 2x6 format for faxes.
(Refer to the M Series Configuration Guide, P/N 9650-0201-01, for more details.)

12-Lead ECG Waveforms

CAUTION

The 12SL analysis results can be affected by poor ECG data quality. If poor data quality is flagged by the system, the interpretation
statements will be preceded by the statement, “Poor data quality, interpretation may be adversely affected.” If this message prints,
the analysis results may be invalid. Check for “LEADS OFF” condition or other sources of noise, correct the condition, and re-
acquire. The ECG waveform should always be reviewed by a physician to confirm any automatic interpretation.

4x3 Format
By default, the 4x3 format prints 10 seconds of 12-lead ECG data in four staggered 2.5 second segments. Each 2.5 second
segment displays the ECG reading for one set of three leads (see figure). The report can be configured for either standard or
Cabrera print formats. 1 mV calibration pulses are printed at the beginning of the report for each data channel. You can also
configure the M Series unit to print 5, 7.5, or 10 seconds of ECG data for each set of leads. If you specify a setting of 5, 7.5, or 10
seconds, the M Series unit prints data for the same time period (5, 7.5, or 10 seconds) for each set of ECG leads; the M Series unit
does not present the ECG data for the 5, 7.5, or 10 second periods in the staggered time format that it uses for the 2.5 second
setting (this option applies only to real time 4x3 printouts, and does not apply to reports or to the data stored to the data card).
For information on how to change this default setting, see the M Series Configuration Guide, P/N 9650-0201-01.

Standard 12-lead printout is always in 4x3 format unless configured differently.

1

st

2.5 sec ECG data 2

nd

2.5 sec ECG data 3

rd

2.5 sec ECG data 4

th

2.5 sec ECG data