OPTION INSERT

9650-0215-01

12-lead - 6

not alter the bandwidth of ECG data passed to the

12SL

program. It only affects the printed waveforms.

Select the filter to be used by pressing the Filter softkey. The
highlight area will scroll among the different filter choices. Press
the Enter softkey to save the highlighted filter and return to the
12 Lead Monitor submenu.

NOTE: When the M Series is turned off for more than 10
seconds, all settings are restored to their original default
settings.

Lead Group Setting

The M Series 12-Lead Option allows any three ECG lead
signals to be simultaneously printed on the strip chart recorder
when using a 12-lead cable. When the Standard lead grouping
is selected, leads I, II and III are printed together, leads aVR,
aVL, and aVF are printed together, leads V1, V2, and V3 are
printed together and leads V4, V5, and V6 are printed together.
When either Custom 1 or Custom 2 lead grouping is selected,
the three (3) leads printed on the recorder will be those
previously programmed for the custom group via the
Configuration mode of the M Series device. (See the
M Series Configuration Guide, P/N 9650-0201-01, for more
information.)

Standard

When programmed to this setting, 3-lead ECG recordings will
be grouped as follows:

leads I, II, and III

leads aVR, aVL, and aVF

leads V1, V2, and V3

leads V4, V5, and V6.

The lead group that prints on the recorder in this mode will be
the group containing the lead selected for display on the
M Series screen. For example, if Lead II is displayed, leads I, II
and III will be printed; if lead V3 is displayed, leads V1, V2, and
V3 will be printed.

NOTE: When the M Series is turned off for more than 10
seconds, all settings are restored to their original default values.

Custom

When the lead group is configured to either Custom 1 or
Custom 2, 3-lead ECG recordings will include the leads
preconfigured for that custom group.

Pressing the Lead Grp softkey will scroll the highlight among
the different selections allowing the user to select the
appropriate Lead Group for three channel printing.

Press the Enter softkey to save the highlighted Lead Group
and return to the 12 Lead Monitor menu.

Dial Type Setting

Selecting the Dial Type setting allows the user to select either
Tone or Pulse dialing.

Pressing the Dial Typ softkey toggles the highlight between
both types of dialing modes. Press the Enter softkey to save
the highlighted dial mode and return to the 12 Lead Monitor
menu. Press the Return softkey to return to the 12 Lead
Monitor menu without saving the highlighted dial mode.

PT Info Softkey

Press the PT INFO softkey to access the patient demographic
information.

Pressing the PT INFO softkey allows the user to select from the
following softkeys: ID#, Age, Gender, Patient Records or
Return.

Note: If no information has been entered, the default age of
“60” will be displayed above the Age softkey and the default
gender of “Male” will be displayed above the Gender softkey.
The 12SL

algorithm processes ECG data based on age. All

patients age 41 and older are handled in the same manner by
the algorithm.

If a patient is under 41 years of age, you must enter the
patient’s age.
This is especially important because the 12SL
algorithm contains age-specific criteria. For patients 15 years
of age or younger, the 12SL algorithm performs a pediatric
analysis.