12-lead

9650-0215-01

12-lead - 7

Patient Identification Number (ID#)

Pressing the ID # softkey causes the “Name and ID#” screen to
appear. “Name” is entered by pressing the Prev Digit and Next
Digit
softkeys to select the digit then pressing the Inc Digit and
Dec Digit softkeys to select the value of the digit. When
“Name” has been entered, press the Enter Name softkey. The
highlight will automatically advance to the “Patient ID#” line.
The Patient ID# is entered exactly the same way as above.
When the ID# has been entered, press the Enter ID and
Return
softkey to return to the Patient Information Menu and
store the name and ID number. If no patient ID # is entered the
M Series automatically generates a 12 digit patient ID# based
on the year/date/time (200002151320) of the first 12-lead
acquisition.

Whenever possible, enter a patient identification number (ID)
before transmitting 12-lead ECG data to a Catalyst MUSE
system. If the data does not have a unique ID number, the
medical facility must search through all of its unidentified
records to find the patient’s information.

Patient Age

Pressing the Age softkey causes the number displayed above
the Age softkey to be highlighted. Press the Dec Age or Inc
Age
softkey to change the patient’s age. Press the Enter
softkey to save the highlighted age and return to the Patient
Information screen. Press the Return softkey to return to the
Patient Information screen without saving the highlighted age.

Patient Gender

Pressing the Gender softkey toggles the highlight between both
genders displayed above the Gender softkey. Press the Enter
softkey to save the highlighted gender and return to the Patient
Information screen. Press the Return softkey to return to the
Patient Information screen without saving the highlighted
gender.

Patient Records

Pressing the Patient Records softkey causes the Patient
Records screen to appear. This screen is used to transmit or
print specific 12-lead ECG records denoted by Patient ID #,
date, and time. (Patient Records are stored in Summary
memory. The user may wish to set Summary

Report Restart

Delay to 1.5 days in order to extend the storage time for 12-
lead Patient Records. See the M Series Configuration Guide,
P/N 9650-0201-01, for instructions on how to implement this
feature.)

Pressing the Prev Record and Next Record softkeys will scroll
the highlight up and down through the list of stored 12-lead
records.

Pressing the Print softkey will print the selected 12-lead patient
record on the strip chart. The format is the same as that stored
in summary memory.

Pressing the Transmit softkey will transmit the selected 12-
lead patient record. The Transmission Setup screen will appear
as described in the Data Transmission section of this manual.

Pressing the Return softkey exits the Patient Records screen
and displays the 12 Lead menu.