Note For monitoring times exceeding 6 hours, Nafion

tubing is recommended.

Selecting a Sidestream Cannula

Select a sidestream cannula based on the patient's size
and monitoring situation. Nasal and nasal/oral cannulae
are disposable and single patient use.

Attaching the LoFlo Module Cable and
Inserting the Sample Cell

Follow these steps:

1. Attach the LoFlo module cable to the CO

2

connector

at the back of the M Series unit.

2. Remove the LoFlo sampling cannula or airway

adapter kit from the package.

3. Insert the LoFlo sample cell into the LoFlo sample

cell receptacle and ensure that it clicks into place.

4. Ensure that the LoFlo module exhaust tube vents

gasses away from the module environment.

5. Turn the selector switch on the M Series to

MONITOR (ON for AED units).

6. Wait for the CO

2

module to warm up.

The unit will display the "WARM UP" message for
approximately one minute while the module and
accessory warm to operating temperature. The
message disappears when the module is ready for
use.

Note Warm up time varies with ambient temperature of

the module.

Zeroing the LoFlo CO

2

Module

The first time a particular LoFlo module is connected to
the M Series, the unit displays the message “ZERO CO2
MODULE” following the warm up period.
The unit retains the LoFlo module zero settings during
and after its use. When the previously used module is
reconnected to the M Series unit, module zeroing does
not have to be repeated, as long as no other LoFlo
module has been connected and zeroed during the
interim period.

Note After the initial zero procedure, the LoFlo module

may be used during the warm up period. The
capnogram, EtCO

2

, and respiratory rate will be

displayed; however, full product specifications are
not achieved until the warm up period has
completed.

CAUTION! Always ensure that the sample cell is

properly connected to the LoFlo module
before zeroing.

1. Ensure that the nasal cannula or on-airway adapter is

not connected to the patient or close to any source of
CO

2

(including the patient's -- and your own --

exhaled breath and ventilator exhaust valves).

Pediatric/Infant Airway Adapter Kit

≤ 4.0 mm

Pediatric/Infant Airway Adapter Kit
with Nafion tubing

Cannula Application

Nasal CO

2

Sampling Cannula, Adult

Nasal CO

2

sampling only

Nasal CO

2

Sampling Cannula,

Pediatric

Nasal CO

2

Sampling Cannula, Infant

Oral/Nasal CO

2

Sampling Cannula,

Adult

Oral/Nasal CO

2

sampling only

Oral/Nasal CO

2

Sampling Cannula,

Pediatric

Nasal CO

2

Sampling with O

2

Delivery

Cannula, Adult

Nasal CO

2

sampling with
oxygen delivery

Nasal CO

2

Sampling with O

2

Delivery

Cannula, Pediatric

Oral/Nasal CO

2

Sampling with O

2

Delivery Cannula, Adult

Oral/Nasal CO

2

sampling with
oxygen delivery

Oral/Nasal CO

2

Sampling with O

2

Delivery Cannula, Pediatric

Airway Adapter Kit

ET Tube Diameter

Figure 4

Module Cable

Exhaust Tube

Sample Cell

Sample Cell
Receptacle

LoFlo Module