Safety considerations, Safety considerations warnings – ZOLL M Series Defibrillator Rev D ETCO2 User Manual

Page 4

Advertising
background image

EtCO

2

- 2

Safety Considerations

WARNINGS

General

Carefully read the M Series Operator’s Guide and
these operating instructions before operating the
EtCO

2

monitoring option.

Ensure that the M Series EtCO

2

option is operated

by qualified personnel only.

Do NOT use the M Series EtCO

2

option as an apnea

monitor.

Do NOT immerse the M Series unit, patient cables or
sensors in water, solvents, or cleaning solutions.

If the accuracy of any reading is suspect, first check
the patient’s vial signs by alternate means and then
check the M Series EtCO

2

option for proper

operation.

If an alarm condition occurs while the alarms are
suspended, the suspended alarm indications will
only be visual displays and symbols. No audio alarm
indications will occur.

To ensure patient safety, ensure that the ECG-out
jack and modem (if available) are only connected to
other equipment with galvanically isolated circuits.

Elevated oxygen levels, nitrous oxide or halogenated
agents contained in the breathing gases may
degrade the accuracy of measurements made with
the M Series EtCO

2

option. Activate oxygen

compensation if O

2

levels in excess of 60% are

introduced. Activate N

2

O compensation if nitrous

oxide is introduced into the airway circuit. The
presence of Desflurane beyond 5% may positively
bias the carbon dioxide reading by up to 3 mmHg.

Do NOT use the LoFlo module on patients who
cannot tolerate the removal of 50ml/min from the
airway.

Carefully route patient cabling and gas sampling
tubes to reduce the possibility of patient
entanglement or strangulation.

Do not touch the bed, patient, or any equipment
connected to the patient during defibrillation. A
severe shock can result. Do not allow exposed
portions of the patient’s body to come in contact with
metal objects, such as a bed frame, as unwanted
pathways for defibrillation current may result.

CAPNOSTAT 3 and LoFlo Sensor and Accessories

Always ensure the integrity of the patient breathing
circuit after insertion of the mainstream or
sidestream airway adapter by verifying a proper CO

2

waveform (capnogram) on the monitor display.

Do NOT use CAPNOSTAT 3 or LoFlo sensors in the
presence of flammable anesthetics or other
flammable gases.

Do not attempt to open the sensor. An electrical
shock hazard exists internally. Refer servicing to
qualified personnel.