Recorder operation, Automated external defibrillator (aed) operation, Semi-automatic operation – ZOLL M Series Defibrillator Rev D ETCO2 User Manual

Page 15: Manual mode operation, Check out procedures, Mainstream etco2 (capnostat 3 sensor), Sidestream etco2 (loflo module), End tidal carbon dioxide (etco

Advertising
background image

End TIdal Carbon Dioxide (EtCO

2

)

EtCO

2

- 13

disabling EtCO

2

or respiration rate alarms disables the

other alarm function. See the M Series Operator’s Guide
for details on enabling, disabling, and suspending alarm
functions on the M Series unit.
When the EtCO

2

and respiration rate alarm states are set

to AUTO, the unit automatically sets the lower and upper
limits for EtCO

2

and respiration rate.

For EtCO

2

, the limits will be set to +/- 25% of the patient's

currently measured EtCO

2

value. If the EtCO

2

value is

greater than 40 mmHg (which is equivalent to 5.3kPa or
5.3% at a barometric pressure of 760 mmHg), then
10 mmHg (1.3 for kPa or %) will be added and
subtracted from the current reading to set the upper and
lower limits. The auto alarm limits are set only if valid
measurements are present for the vital sign.
For the automatic respiration rate alarm limits, the unit
sets the upper and lower limits for respiration by adding
and subtracting the values shown in the following table
to/from the patient's current breath rate.

Recorder Operation

If EtCO

2

measurements have been taken, press the

RECORDER button to print a stripchart that includes the
following values across the top part of the paper:

date and time

ECG lead and size

heart rate

EtCO

2

value

respiration rate

The recorder runs continuously until the button is
pressed again. If displayed on the screen, the
capnogram waveform will also be printed under the ECG
trace. All waveforms printed by the recorder are delayed
by six seconds relative to their occurrence.

Automated External Defibrillator (AED)
Operation

M Series AED units equipped with the EtCO

2

option

operate in a slightly different way than Manual and
Advisory models equipped with EtCO

2

, as outlined

below.

Semi-Automatic Operation

The EtCO

2

monitoring parameters can be changed by

pressing the Param softkey, as outlined in "Physiological

Monitoring" on page 11. The capnogram waveform
cannot be displayed in semi-automatic mode.
Although EtCO

2

alarm functions are operational in

semi-automatic mode, heart rate alarm functions are
disabled. Background ECG analysis functions continue
to operate as described in the “AED” section of the
M Series Operator’s Guide.
The ALARM SUSPEND button can be used to activate,
deactivate, or audibly disable the EtCO

2

alarms (as

described in the M Series Operator’s Guide). The alarm
limit settings cannot, however, be changed in
semi-automatic mode; only the default alarm limits are
available. See the M Series Configuration Guide for
information on setting alarm limit defaults.

Manual Mode Operation

When the AED unit is in manual mode, the unit can
display the capnogram waveform as described in
"Displaying the Capnogram Waveform" on page 11.
Both heart rate and EtCO

2

alarms are operational. The

alarm limits can be changed by pressing the Alarms
softkey. The EtCO

2

monitoring parameters can be

changed by pressing the Param softkey, as outlined in
"Physiological Monitoring" on page 11.

Check Out Procedures

The following procedures verify that the EtCO

2

option is

functioning properly.

Mainstream EtCO

2

(CAPNOSTAT 3 Sensor)

1. Connect the CAPNOSTAT 3 sensor cable to the

EtCO

2

connector at the back of the M Series unit.

2. Connect an airway adapter to the sensor.
3. Turn the selector switch to MONITOR mode (ON for

AED units and select Manual Mode).

4. Wait for the CO

2

sensor to warm up. The message

"WARM UP" is displayed for approximately one
minute.

5. Perform a zero procedure if necessary (see "Zeroing

the Mainstream Sensor/Airway Adapter" on page 6).

6. Breath normally into the adapter.
7. Verify that the unit displays EtCO

2

readings in the

EtCO

2

display area of the monitor.

8. Verify the capnogram waveform is displayed by

pressing the Wave2 softkey.

Sidestream EtCO

2

(LoFlo Module)

Use an Adult/Pediatric Airway Adapter when performing
this procedure.

1. Connect the LoFlo module cable to the EtCO

2

connector at the back of the M Series unit.

2. Insert the adapter sample cell into the LoFlo module

sample cell receptacle.

3. Turn the selector switch to MONITOR mode (ON for

AED units and select Manual Mode), and wait

Respiration Limits (Auto)

Respiration Rate

Average

High Limit

Low Limit

1-15 breaths/min.

+7 breaths/min.

-50% value

16-40 breaths/min.

+10 breaths/min.

-7 breaths/min.

> 40 breaths/min.

+15 breaths/min.

-10 breaths/min.

Advertising