Specifications

This section summarizes the specifications of the M Series End-Tidal Carbon Dioxide (EtCO

2

) option.

CAPNOSTAT 3

LoFlo

Transducer Type

Mainstream

Sidestream

Principle of Operation

Non-Dispersive Infrared (NDIR) single beam optics, dual wavelength, no moving
parts.

Warm Up Time

Full specifications within 60 seconds. Capnogram in 15 seconds.

EtCO

2

Measurement Range

0 - 100 mmHg
0 - 13%
0 - 12.5 kPa

EtCO

2

Accuracy

(at 760 mmHg, ambient
temperature of 25°C)

0 - 40 mmHg, ±2 mmHg
41 - 70 mmHg, ±5% of actual
71 - 100 mmHg, ±8% of actual

0 - 40 mmHg, ±2 mmHg
41 - 70 mmHg, ±5% of actual
71 - 100 mmHg, ±8% of actual
±12% for respiration rate above
80 BrPM

EtCO

2

Resolution

0.5 mmHg

EtCO

2

Stability

Short Term Drift: Drift over four hours < 0.8 mmHg.
Long Term Drift: Accuracy specification will be maintained over a 120 hour period
after zeroing.

EtCO

2

Noise

RMS noise of the sensor < 0.25 mmHg at 7.5% CO

2

.

EtCO

2

Rise Time (10-90%)

< 60 ms (Adult/pediatric adapters)
< 50 ms (Neonatal adapters)

< 200 ms

Respiration Rate Range

0 - 150 breaths per minute

Respiration Rate Accuracy

±1 breath per minute

Sample Flow Rate

N/A

50 ml/min

Compensations

Barometric pressure 550 - 780 mmHg (automatic).
Operator selectable O

2

/N

2

O compensation.

EtCO

2

Alarm Limits

User selectable, High 5 - 100 mmHg, Low 0 - 95 mmHg, OFF.

Respiration Rate (RR) Alarm
Limits

User selectable, High 5 - 150 respirations per minute, Low 0 - 100 respirations per
minute, OFF.

Halogenated Agents

Specification allows for halogenated anesthetic agents that may be present at
normal clinical levels. The presence of Desflurane in the exhaled breath beyond
normal values (5-6%) may positively bias carbon dioxide values by up to an
additional 2-3 mmHg.

Airway Adapter Deadspace

Adult < 5 cc
Infant < 1.0 cc

Adult 7.3 cc maximum
Pediatric/Infant < 1 cc

Environmental

Operating Temperature:

10º C to 40º C

Storage and Shipping Temperature: -10º C to 55º C
The M Series unit may not perform to specifications when stored at the upper or
lower extreme limits of storage temperature and immediately put into use.

Electromagnetic Immunity

AAMI DF-39, IEC 60601-1-2, 20 V/m