12-Lead ECG Monitoring Setup

9650-0806-01 Rev. C

Propaq MD Operator’s Guide

14-11

You can then view four 12-Lead snapshot pages by pressing

to step through them

sequentially. For example, the unit displays the following 12-Lead snapshot after the Analysis
Page 2
:

Fault Conditions Affecting 12-Lead Interpretive Analysis

The Propaq MD unit will not perform Interpretive Analysis if it detects one of the following
fault conditions when attempting to acquire 12-Lead data:

A pacer signal is present.

A lead fault is detected in the ECG cable.

An invalid cable is in use.

If the Propaq MD unit detects one of these fault conditions, Analysis Page 1 indicates that there
is No data available for Interpretive Analysis and lists the fault condition; all measurements on
Analysis Page 2 appear as N/A.

Once you have corrected the fault condition, press

to confirm the correction and perform

the 12-Lead Interpretive Analysis.