background image

Transferring Data to a USB Device

9650-0806-01 Rev. C

Propaq MD Operator’s Guide

21–5

Warning!

To avoid a possible shock hazard, do NOT make any electrical connections to the USB
port except to connect a USB flash drive while in the vicinity of the patient.

Clearing the Log

You should clear the patient data log after transferring data to the USB device or when the log is
full.

Note:

Clearing the log during patient treatment results in the loss of all patient data and
events recorded prior to clearing the log. Clearing the log creates a New Patient record
and all patient-specific parameters (alarm limit, defibrillation energy, etc.) are set to
their default values.

To clear the log:

1. Press .

2.

Press .

3.

Press the Clear Log quick access key (

).

3. Use the navigation keys to select Yes.

Note:

An UNABLE TO READ LOG message indicates that the log contains no information.
This message can occur if you clear the log and then immediately enter the Treatment
screen or the Trend Summary screen.

LOG