Defibrillator output energy, External pacemaker, Ecg monitoring – ZOLL Propaq MD Rev C User Manual

Page 23: Electrodes

Advertising
background image

Propaq MD Product Functions

9650-0806-01 Rev. C

Propaq MD Operator’s Guide

1-11

Defibrillator Output Energy

Propaq MD defibrillators can deliver biphasic energy from 1 joule to 200 joules. The energy
delivered through the chest wall, however, is determined by the patient’s transthoracic
impedance. An adequate amount of electrolyte gel must be applied to the paddles and a force of
10 to 12 kilograms (22 to 26.4 pounds) must be applied to each paddle in order to minimize this
impedance. If hands-free therapy electrodes are used, make sure that they are properly applied.
(Refer to the instructions on the electrode package).

External Pacemaker

Propaq MD defibrillators include a transcutaneous pacemaker consisting of a pulse generator
and ECG-sensing circuitry. Noninvasive transcutaneous pacing (NTP) is an established and
proven technique. This therapy is easily and rapidly applied in both emergency and
nonemergency situations when temporary cardiac stimulation is indicated.

The output current of the pacemaker is continuously variable from 10 to 140 mA (the output
current is 0 mA when paused). The rate is continuously variable from 30 to 180 pulses per
minute (ppm), by increments of 5 ppm (10 bpm when greater than 100 ppm).

The pacing output pulse is delivered to the heart via ZOLL hands-free defibrillation/pacing
electrodes placed on the patient’s back and the precordium.

Proper operation of the device, together with correct electrode placement, is critical to
obtaining optimal results. Every operator must be thoroughly familiar with these operating
instructions.

ECG Monitoring

The patient’s ECG is monitored by connecting the patient to the unit via a 3-, 5-, or 12-lead
patient cable or hands-free therapy electrodes. The ECG waveform is presented on the display
along with the following information:

averaged heart rate, derived by measuring R to R intervals

lead selection - I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 (with ECG cable),
PADDLES, or PADS.

ECG size - 0.125, 0.25, 0.50, 1.0, 2.0, 4.0 cm/mV, AUTO

status messages

The ECG bandwidth is user selectable.

Electrodes

The Propaq MD units will defibrillate, cardiovert, and monitor ECG using hands-free therapy
electrodes. The Propaq MD unit will pace using ZOLL hands-free therapy electrodes.

Energy Select, Charge and Shock controls are located on the paddles and front panel. When
using hands-free therapy electrodes, you must use the controls on the front panel of the unit. To
switch between paddles and hands-free therapy electrodes, remove the multifunction cable
(MFC or OneStep) from the apex paddle and connect the hands-free therapy electrodes to the
cable.