Setting Alarms

The R Series NIBP option includes blood pressure alarms for the measurements shown in the
following table.

Before taking a measurement, ensure that all the NIBP alarms settings are appropriate for the
situation. See the R Series Operator’s Guide for instructions on setting alarms. When an NIBP
or heart rate alarm is triggered, the unit automatically initiates a single blood pressure
measurement. See the R Series Configuration Guide for instructions on how to enable/disable
this automatic measurement.

When the R Series unit is turned on, all alarm functions are disabled, and the alarm limits are
set to their default values for the selected patient type. Any changes to these settings remain in
effect until either the settings are changed for the patient type or for 10 seconds after the
R Series unit is turned off. You can configure the default alarm limits at power-up; see the
R Series Configuration Guide for more information.

Setting Alarm Limits

You can change all the alarm limits using standard R Series procedures. See the R Series
Operator’s Guide
for instructions on setting alarm limits.

Patient Type

Alarm Parameter

Default Setting

Range (increments of 5)

Adult

Systolic

High

160 mmHg (21.3 kPa)

80-260 mmHg (10.7-34.7 kPa)

Low

90 mmHg (12.0 kPa)

40-140 mmHg (5.3-18.7 kPa)

Diastolic

High

110 mmHg (14.7 kPa)

50-200 mmHg (6.7-26.7 kPa)

Low

50 mmHg (6.7 kPa)

25-100 mmHg (3.3-13.3 kPa)

Mean

High

130 mmHg (17.3 kPa)

60-220 mmHg (8.0-29.3 kPa)

Low

60 mmHg (8.0 kPa)

30-120 mmHg (4.0-16.0 kPa)

Pediatric

Systolic

High

145 mmHg (19.3 kPa)

80-160 mmHg (10.7-21.3 kPa)

Low

75 mmHg (10.0 kPa)

35-140 mmHg (4.7-18.7 kPa)

Diastolic

High

100 mmHg (13.3 kPa)

50-130 mmHg (6.7-17.3 kPa)

Low

35 mmHg (4.7 kPa)

20-100 mmHg (2.7-13.3 kPa)

Mean

High

110 mmHg (14.7 kPa)

60-140 mmHg (8.0-18.7 kPa)

Low

50 mmHg (6.7 kPa)

30-120 mmHg (4.0-16.0 kPa)

Neonate

Systolic

High

100 mmHg (13.3 kPa)

60-130 mmHg (8.0-17.3 kPa)

Low

50 mmHg (6.7 kPa)

25-120 mmHg (3.3-16.0 kPa)

Diastolic

High

70 mmHg (9.3 kPa)

30-105 mmHg (4.0-14.0 kPa)

Low

30 mmHg (4.0 kPa)

20-100 mmHg (2.7-13.3 kPa)

Mean

High

80 mmHg (10.7 kPa)

35-110 mmHg (4.7-14.7 kPa)

Low

35 mmHg (4.7 kPa)

30-105 mmHg (4.0-14.0 kPa)