WARNING!

Make sure that a measurement has already started before attempting to abort. If you
press the NIBP button while a measurement is not being taken, the R Series unit starts a
new measurement.

Note that:

If you press the NIBP button while taking a STAT measurement, the R Series unit
terminates all subsequent measurements.

If you press the NIBP button while the R Series unit is taking an automatic measurement, it
aborts the current measurement, deflates the cuff, and takes all subsequent measurements at
the set interval.

Viewing Trend Data

To view the patient’s trend data over the past 4 hours, press the Trend softkey. The time,
HR/PR, and readings will be displayed.

Measurements with alarm values are highlighted; an asterisk (*) shows measurements that
contain artifact.

3

1

SHOCK

SHOCK

2

ENERGY

SELECT

ENERGY

SELECT

ANALYZE

ANALYZE

CHARGE

CHARGE

OUTPUT

mA

OUTPUT

mA

RATE

ppm

RATE

ppm

4:1

LEAD

SIZE

ALARM

SUSPEND

RECORDER

DEFIB

OFF

MONITOR

PACER

Options

Param

Code

Marker

Report

Data

Alarms

11:24

72

ECG

PADS

x1

SpO2%

CPR

0:00

IDLE

Depth
Rate

99

MONITOR

NIBP

ECG

mmHg

---
---

---

CO2

RR

Ɔ

---
---

mmHg

NIBP Button

Return

TIME

HR/PR

NIBP mmHg

SYS DIA M

11:22

92

127 81 96

Newer

Older

Print

11:27

11:32

11:37

11:42

11:47

11:52

11:57

12:02

91

91

92

72

86

87

91

92

127 81 96

126 72 99

126 73 100

125 72 99

127 81 96

88

54 65

86

65

72 *

81

53 62

SpO2

%

99

99

98

99

98

99

97

98

99

Advertising