<8. Maintenance>

8-4

IM 01C25B01-01E

8.4.2 Replacing the CPU Board Assembly

This subsection describes the procedure for

replacing the CPU assembly. (See figure 8.3)

■ Removing the CPU Assembly
1) Remove the cover. If an integral indicator is

mounted, refer to subsection 8.4.1 and remove

the indicator.

2) Turn the zero-adjustment screw to the position

(where the screw head slot is horizontal) as

shown in figure 8.3.

3) Disconnect the output terminal cable (cable

with brown connector at the end). When doing

this, lightly press the side of the CPU assembly

connector and pull the cable connector to

disengage.

4) Use a socket driver (width across flats, 5.5mm)

to loosen the two bosses.

5) Carefully pull the CPU assembly straight

forward to remove it.

6) Disconnect the flat cable (cable with white

connector at the end) that connects the CPU

assembly and the capsule.

NOTE

Be careful not to apply excessive force to the

CPU assembly when removing it.

■ Mounting the CPU Assembly
1) Connect the flat cable (with white connector)

between the CPU assembly and the capsule.

2) Connect the output terminal cable (with brown

connector).

NOTE

Make certain that the cables do not get pinched

between the case and the edge of the CPU

assembly.

3) Align and engage the zero-adjustment screw

pin with the groove on the bracket on the CPU

assembly. Then insert the CPU board assembly

straight onto the post in the amplifier case.

4) Tighten the two bosses. If the transmitter is

equipped with an integral indicator, refer to

subsection 8.4.1 to mount the indicator.

NOTE

Confirm that the zero-adjustment screw pin is

placed properly in the groove on the bracket prior

to tightening the two bosses. If it is not, the zero-

adjustment mechanism will be damaged.

5) Replace the cover.

8.4.3 Cleaning and Replacing the Capsule

Assembly

This subsection describes the procedures for

cleaning and replacing the capsule assembly. (See

figure 8.4.)

CAUTION

Cautions for Flameproof Type Transmitters

Users are prohibited by law from modifying the

construction of a flameproof type transmitter. If

you wish to replace the capsule assembly with

one of a different measurement range, contact

Yokogawa.

The user is permitted, however, to replace a

capsule assembly with another of the same

measurement range. When doing so, be sure to

observe the following.

• The replacement capsule assembly must

have the same part number as the one being

replaced.

• The section connecting the transmitter and

capsule assembly is a critical element in

preservation of flameproof performance, and

must be checked to verify that it is free of

dents, scratches, and other defects.

• After completing maintenance, be sure to

securely tighten the setscrews that fasten

the transmitter section and pressure-detector

section together.