8.5.3 Alarms and Countermeasures

Table 8.3

Alarm Message Summary

Indicator

Cause

Output Operation

during Error

Countermeasure

None
AL. 01

CAP. ERR

Sensor problem.

Outputs the signal (Hold, High, or

Low) set with parameter.

Replace capsule when error

keeps appearing even after

restart.

Capsule temperature sensor

problem.
Capsule EEPROM problem.

AL. 02

AMP. ERR

Amplifier temperature sensor

problem.

Outputs the signal (Hold, High, or

Low) set with parameter.

Replace amplifier.

Amplifier EEPROM problem.
Amplifier problem.

AL. 10

PRESS

Input is outside measurement

range limit of capsule.

Outputs high range limit value or

low range limit value.

Check input or replace capsule

when necessary.

AL. 11

ST. PRSS

Static pressure exceeds limit.

Continues to operate and output.

AL. 12

CAP. TMP

Capsule temperature is outside

range (–50 to 130°C).

Use heat insulation or make

lagging to keep temperature

within range.

AL. 13

AMP. TMP

Amplifier temperature is outside

range (–50 to 95°C)

AL. 30

RANGE

Output is outside upper or lower

range limit value.

Outputs high range limit value or

low range limit value.

Check input and range setting,

and change them as needed.

AL. 31

SP. RNG

Static pressure exceeds specified

range.

Holds output immediately before

error occurred.

AL. 35 *

1

P. HI

Input pressure exceeds specified

threshold.

Continues to operate and

output.

Check input.

AL. 36 *

1

P. LO
AL. 37 *

1

SP. HI

Input static pressure exceeds

specified threshold.

AL. 38 *

1

SP. LO
AL. 39 *

1

TMP. HI

Detected temperature exceeds

specified threshold.

AL. 40 *

1

TMP. LO
AL. 50

P. LRV

Specified value is outside of setting

range.

Holds output immediately before

error occurred.

Check setting and change them

as needed.

AL. 51

P. URV
AL. 52

P. SPN
AL. 53

P. ADJ

Continues to operate and output. Check input.

AL. 54

SP. RNG

Continues to operate and output

holding static pressure in %.

Check setting and change them

as needed.

AL. 55

SP. ADJ

Continues to operate and output. Check input.

AL. 60

SC. CFG

Specified values or settings to

define signal characterizer function

do not satisfy the condition.

Check setting and change

them as needed.

AL. 79

OV. DISP

Displayed value exceeds limit.