Yokogawa EJA110E User Manual

Page 52

Advertising
background image

<9. General Specifications>

9-4

IM 01C25B01-01E

Maximum Over Pressure

EJ310

Capsule

Pressure

L, M

500 kPa abs (72 psia)

A

16 MPa abs (2300 psia)

B

25 MPa abs (3600 psia)

EJ430

Capsule

Pressure

H, A

16 MPa (2300 psi)

B

25 MPa (3600 psi)*

* 24 MPa for wetted parts material code H, M, T, A, D, and

B.

EJ440

Capsule

Pressure

C

48 MPa (6750 psi)

D

75 MPa (10800 psi)*

* 60 MPa (8700 psi) for EJA440E.

Supply & Load Requirements “◊” (Optional

features or approval codes may affect

electrical requirements.)

With 24 V DC supply, up to a 550 Ω load can be

used. See graph below.

E-10.5

0.0244

(Ω)

Power supply voltage E (V DC)

600

250

R

10.5

16.6

25.2

42

External

load

resistance

Digital

Communication

range

BRAIN and HART

R=

F0903.ai

Figure 9.2

Relationship Between Power Supply

Voltage and External Load Resistance

Supply Voltage

10.5 to 42 V DC for general use and flameproof

type.

10.5 to 32 V DC for lightning protector (Option

code /A).

10.5 to 30 V DC for intrinsically safe, type n or

non-incendive type.

Minimum voltage limited at 16.6 V DC for digital

communications, BRAIN and HART

Load (Output signal code D, E and J)

0 to 1290 Ω for operation

250 to 600 Ω for digital communication

Communication Requirements “◊”

(Safety approvals may affect electrical

requirements.)

BRAIN

Communication Distance

Up to 2 km (1.25 miles) when using CEV

polyethylene-insulated PVC-sheathed cables.

Communication distance varies depending on

type of cable used.
Load Capacitance

0.22 μF or less
Load Inductance

3.3 mH or less
Input Impedance of communicating device

10 kΩ or more at 2.4 kHz.

HART

Communication Distance

Up to 1.5 km (1 mile) when using multiple

twisted pair cables. Communication distance

varies depending on type of cable used.

EMC Conformity Standards

,

EN61326-1 Class A, Table2 (For use in industrial

locations)

EN61326-2-3