<9. General Specifications>

9-20

IM 01C25B01-01E

Item

Description

Code

Oil-prohibited use with

dehydrating treatment *

4

Degrease cleansing treatment and dehydrating treatment.

K5

Degrease cleansing treatment and dehydrating treatment with fluorinated oilfilled

capsule. Operating temperature –20 to 80°C( –4 to 176°F)

K6

Capsule fill fluid

Fluorinated oil filled in capsule

Operating temperature –20 to 80°C( –4 to 176°F)

K3

Calibration units *

5

P calibration (psi unit)

(See Table for Span and

Range Limits.)

D1

bar calibration (bar unit)

D3

M calibration (kgf/cm

2

unit)

D4

Long vent *

6

Total length: 119 mm (standard: 34 mm); Total length when combining with

Optional code K1, K2, K5, and K6: 130 mm. Material: 316SST.

U1

Gold-plated capsule gasket

*

13

Gold-plated 316L SST capsule gasket. No PTFE is used for wetted parts.

GS

Gold-plated diaphragm *

14

Surface of isolating diaphragm is gold plated, effective for hydrogen permeation.

A1

130 Pa abs calibration *

12

Minimum input puressure at calibration testing: 130 Pa abs (1 mmHg abs)

S1

Output limits and failure

operation *

7

Failure alarm down-scale: Output status at CPU failure and hardware error is

–5%, 3.2 mA DC or less.

C1

NAMUR NE43 Compliant

Output signal limits:

3.8 mA to 20.5 mA

Failure alarm down-scale: Output status at CPU

failure and hardware error is –5%, 3.2 mA DC or less.

C2

Failure alarm up-scale: Output status at CPU

failure and hardware error is 110%, 21.6 mA or more.

C3

Body option *

8

Without drain and vent plugs

N1

N1 and Process connection, based on IEC61518 with female thread on both sides

of cover flange, with blind kidney flanges on back.

N2

N2 and Material certificate for cover flange, diaphragm, capsule body, and blind

kidney flange

N3

Stainless steel tag plate

304SST tag plate wired onto transmitter

N4

Data configuration at

factory *

9

Data configuration for HART

communication type

Software damping, Descriptor, Message

CA

Data configuration for BRAIN

communication type

Software damping

CB

European Pressure

Directive *

10

PED 97/23/EC

Category: III, Module: H, Type of Equipment; Pressure Accessory-Vessel, Type of

Fluid; Liquid and gas, Group of Fluid: 1 and 2

Lower limit of Process and Ambient temperature for EJX110A : –29°C

PE3

Advanced diagnostics *

27

Multi-sensing process monitoring

• Impulse line blockage detection *

28

• Heat trace monitoring

DG6

Material certificate *

15

Cover flange *

16

M01

Cover flange, Process connector *

17

M11

Pressure test/

Leak test certificate

Test Pressure: 16 MPa (2300 psi) *

18

Nitrogen(N

2

) Gas *

25

Retention time: one minute

T12

Test Pressure: 25 MPa (3600 psi) *

19

T13

Test Pressure: 3.5 MPa (500 psi)*

20

T01

Test Pressure: 500 kPa (2000 inH

2

O) *

21

T11

Test Pressure: 50 kPa (200 inH

2

O) *

22

T04

Test Pressure: 32 MPa (4500 psi) *

23

Nitrogen(N

2

) Gas or water *

26

Retention time: one minute

T09

Test Pressure: 50 MPa (7200 psi) *

24

T08

High accuracy type

High accuracy

HAC

High pressure-proof

structure

Maximum working pressure for differential pressure measurement: 25MPa*

29

HG

Contact Yokogawa representative for the codes indicated as ‘-’.

*1:

Applicable for Electrical connection code 2, 4, 7, and 9.

*2:

Applicable for Electrical connection code 2 and 7.

*3:

Not applicable for option code /AL.

*4:

Applicable for Wetted parts material code S, H, M, and T.

*5:

The unit of MWP (Max. working pressure) on the name plate of the housing is the same unit as specified by Option code D1, D3,

and D4.

*6:

Applicable for vertical impulse piping type (Installation code 7) and Wetted parts material code S, H, M, and T.

*7:

Applicable for output signal code D, E and J. The hardware error indicates faulty amplifier or capsule.

*8:

Applicable for Wetted parts material code S, H, M, and T; Process connection code 3, 4, and 5; Installation code 9; and Mounting

bracket code N. Process connection faces on the other side of zero adjustment screw.