2 zero point adjustment, Zero point adjustment -2, Important – Yokogawa EJX118A User Manual

Page 32

Advertising
background image

<6. Operation>

6-2

IM 01C25H01-01E

Using the integral indicator

• If the wiring system is faulty, the display stays

blank.

• If the transmitter is faulty, an error code is

displayed.

Self-diagnostic error on the integral indicator

(Faulty transmitter)

F0604.ai

NOTE

If any of the above errors are indicated on

the display of the integral indicator or the

communicator, refer to subsection 7.5.3 for

corrective action.

■ Verify and Change Transmitter Parameter

Setting and Values

The parameters related to the following items are

set at factory as specified by the customer.

• Calibration range

• Integral indicator display

• Output mode

• Software damping (optional)

Other parameters like following are shipped with the

default setting.

• Low-cut

• Process alarm setting

• Static pressure range

• Signal characterizer

• Write protection

To confirm or change the values, see IM 01C25T01-

06EN or 01C25T03-01E.

6.2 Zero Point Adjustment

After completing preparations for operating the

transmitter, adjust the zero point.
Zero point adjustment can be done by turning the

transmitter’s zero-adjustment screw or by using

the communicator. This section describes the

procedure for the zero-adjustment screw. For the

communicator procedure, see the communication

manual.

IMPORTANT

Do not turn off the power to the transmitter

immediately after performing a zero point

adjustment. Powering off within 30 seconds of

performing this procedure will return the zero

point to its previous setting.

NOTE

Before performing this adjustment, make sure

that the external zero adjustment function has

NOT been disabled by a parameter setting.

To check the output signal, use a digital multimeter,

calibrator, or communicator.

(1) When you can obtain the Low Range Value

from the actual measured value of 0%

(0 kPa, atmospheric pressure);

Zero-adjustment

screw cover

F0605.ai

The zero-adjustment screw is located inside the

cover.
Use a slotted screwdriver to turn the zero-

adjustment screw. Turn the screw clockwise

to increase the output or counterclockwise to

decrease the output. The zero point adjustment can

be made with a resolution of 0.01% of the setting

range. The degree of zero adjustments varies with

the screw turning speed; turn the screw slowly to

make a fine adjustment, quickly to make a rough

adjustment.
When adjusting the transmitter zero point, the

liquid level in a tank does not have to be set to

the low limit (0%) of the measuring range; use a

digital manometer or a glass gauge to match the

transmitter output signal with the actual measured

value.

Advertising