<8. General Specifications>

8-9

IM 01C25H01-01E

The “►” marks indicate the most typical selection for each specification. Example: EJX118A-DMSCG-912EN-WA13B1SW00-BA25/

*1:

See table 8.3 ‘Gasket contact surface’ on page 8-4.

*2:

When specified flushing connection ring code 1, 2, A, or B, exclusive gasket is provided for transmitter side.

*3:

In case of wetted parts material code TW (Tantalum), the process temperature limit is –10 to 200°C.

*4:

Wetted parts material code TW (Tantalum) cannot be applied.

*5:

Even in case where fill fluid code D (fluorinated oil) is selected, if degrease cleansing treatment or both degrease cleansing and

dehydrating treatment for the wetted parts is required, specify option code K1 or K5.

*6:

In case of wetted parts material code HW (Hastelloy C), TW (Tantalum), and UW (Titanuym) for 2-inch pressure flange, specify

capillary length from 1 to 5 m.

*7:

Flushing connection ring code 1, 2, A, or B cannot be applied.

*8:

Not applicable for gasket contact surface code 1.

*9:

Hastelloy C-276 or N10276.

*10: Forged version of the material may be used.

*11: Not applicable for wetted parts material code UW (Titanium).

*12: The upper ambient temperature limit is 50°C (122°F) in the following combinations.

Process connection style code

Process connection size code

W (Flush type)

2 (2-inch) or 8 (1 1/2-inch)

E (Extension type)

3 (3-inch)

The “#” marks indicate the construction materials conform to NACE material recommendations per MR01-75. For the use of 316 SST

material, there may be certain limitations for pressure and temperature. Please refer to NACE standards for details.