<8. Maintenance>

7-2

IM 01C25H01-01E

Table 7.1

Instruments Required for Calibration

Name

Yokogawa-recommended Instrument

Remarks

Power supply Model SDBT or SDBS distributor

4 to 20 mA DC signal

Load resistor

Model 2792 standard resistor [250 Ω ±0.005%, 3 W]
Load adjustment resistor [100 Ω ±1%, 1 W]

Voltmeter

Model 2501 A digital multimeter

Accuracy (10V DC range): ±(0.002% of rdg + 1 dgt)

Digital

manometer

Model MT220 precision digital manometer

1) For 10 kPa class

Accuracy: ±(0.015% of rdg + 0.015% of F.S.). . . . for 0 to 10 kPa

±(0.2% of rdg + 0.1% of F.S.). . . . . . . . for -10 to 0 kPa

2) For 130 kPa class

Accuracy: ±0.02% of rdg. . . . . . . . . . . . . . . . . . . . for 25 to 130 kPa

±5digits. . . . . . . . . . . . . . . . . . . . . . . . . for 0 to 25 kPa

±(0.2% of rdg + 0.1% of F.S.). . . . . . . . for -80 to 0 kPa

3) For 700 kPa class

Accuracy: ±(0.02% of rdg + 3digits) . . . . . . . . . . . for 100 to 700 kPa

±5 digits . . . . . . . . . . . . . . . . . . . . . . . . for 0 to 100 kPa

±(0.2% of rdg + 0.1% of F.S.). . . . . . . . for -80 to 0 kPa

4) For 3000 kPa class

Accuracy: ±(0.02% of rdg + 10 digits). . . . . . . . . . for 0 to 3000 kPa

±(0.2% of rdg + 0.1% of F.S.). . . . . . . . for -80 to 0 kPa

5) For 130 kPa abs class

Accuracy: ±(0.03% of rdg + 6 digits) . . . . . . . . . . . for 0 to 130 kPa abs

Select a manometer having

a pressure range close to

that of the transmitter.

Pressure

generator

Model 7674 pneumatic pressure standard for 200 kPa {2 kgf/cm

2

}, 25 kPa

{2500 mmH

2

O}

Accuracy: ±0.05% of F.S.

Requires air pressure

supply.

Dead weight gauge tester 25 kPa {2500 mmH

2

O}

Accuracy: ±0.03% of setting

Select the one having a

pressure range close to that

of the transmitter.

Pressure

source

Model 6919 pressure regulator (pressure pump)

Pressure range: 0 to 133 kPa {1000 mmHg}

Prepare the vacuum pump

for negative pressure

ranges.

Note: The above table contains the instruments capable of performing calibration to the 0.2% level. Since special maintenance and

management procedures involving traceability of each instrument to higher-level standards are required for calibration to the 0.1%

or higher level, there may be difficulties in calibration to this level in the field. For calibration to the 0.1% level, contact Yokogawa

representatives from which the instrument was purchased or the nearest Yokogawa office.

P

P

V

Mating calibration flange

Load resistance, 250Ω

Model MT220

precision digital manometer

High pressure side

diaphragm seal

Rc

R

Load adjustment

resistance, 100Ω

Digital voltmeter

Power

supply

E

Low pressure side diaphragm seal

Model 7674 pneumatic

pressure standards

Pressure source

Supply pressure

If a pressure

generator is used:

If a pressure source

and a manometer

are combined:

F0701.ai

Figure 7.1

Instrument Connections