3-1

3. COMPONENT NAMES

3.

COMPONENT NAMES

F0301.EPS

Vertical impulse piping type

Pressure-detector section

Cover
flange

Process connection

Process connector

(Note 1)

External indicator
conduit connection

(Note 1)

Terminal box cover

Horizontal impulse piping type

External indicator
conduit connection

(Note 1)

Conduit
connection

Conduit
connection

Zero-
adjustment
screw

Cover
flange

Process connection

Process connector

(Note 1)

Transmitter section

Setting pin (CN4)

Integral
indicator

(Note 1)

CPU assembly

Mounting screw

Range-setting
switch

(Note 1)

(See Subsection 7.6)

Amplifier Cover

Setting pin (CN4)
position

(Note 2)

Burn - out
direction

Output at
burn - out

H

L

H

L

HIGH

LOW

110% or

higher

-5% or

lower

Open to atmosphere
(∅5mm)

Note 1: See Subsection 10.2, “Model and Suffix Codes,” for

details.

Note 2: Insert the pin (CN4) as shown in the figure above to

set the burn-out direction. The pin is set to the H side
for delivery (unless option code /C1 is specified in the
order).
The setting can be confirmed by calling up parameter
D52 using the BRAIN TERMINAL. Refer to Subsection
8.3.3 (8).

Figure 3.1 Component Names

Table 3.1 Display Symbol

Display Symbol

%, Pa, kPa, MPa, kgf/cm

2

, gf/cm

2

, mbar, bar,

atm, mmHg, mmH

2

O, inH

2

O, inHg, ftH

2

O, psi, Torr

Meaning of Display Symbol

The output signal being zero-adjusted is increasing.

The output signal being zero-adjusted is decreasing.

Select one of these sixteen available engineering units for the display.